SPECIFICATIONS
Sample:
Whole Blood, Serum, Plasma, or Urine
Sample Size:
100 µl Whole Blood, Serum or Plasma
400 µl diluted (1:10) Urine
Detection Range: Blood Na+: 45-205 mmol/L

K+: 1.5-11 mmol/L

Cl-: 45-205 mmol/L

Urine Na+: 30-1020 mmol/L (undiluted)

K+: 20-505 mmol/L (undiluted)

Cl-: 25-505 mmol/L (undiluted)
Analysis Time: 55 sec (Blood)
90 sec (Urine)
Data Storage: 125 patient results
20 normal QC results
40 abnormal QC results
Calibration: Automatic or On Demand
Reproducibility: Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV ≤ 1% (80-200 mmol/L)
K+: CV ≤ 2% (1-10 mmol/L)
Cl-: CV ≤ 2% (80-200 mmol/L)
Urine
Na+: CV ≤ 5% (25-1000 mmol/L)
K+: CV ≤ 5% (1-500 mmol/L)
Cl-: CV ≤ 5% (25-1000 mmol/L) |
Output: 16x2 place alphanumeric display
40 column printer
Serial port (RS-232)
AutoSampler port
Ambient Conditions: 15-32°C (60-90°F), <85% humidity
Power: 110-115 ~ VAC 50-60 Hz, 0.8 A or
220~VAC 50/60 Hz, 0.4 A
(Factory set)
Size & Weight: 9.5” (24cm) W x 16.5” (42 cm) H x
8.0” (20cm) D, 13 lbs. (5.8 kg) |